Femoral introducer sets Radial introducer sets
|
[Product Structure] The product is composed of introducer sheath, dilator, guide wire, and seldinger needle. The product is made from polyethylene, polypropylene, and stainless steel.
[Product Type] 5F, 6F, 7F, 8F
[Application Scope] The introducer set and dilator enter into the vas along guidewire to build a passage, which connects skin and vas. After the dilator pulling out, the introducer sheath is the only entrance for the vas operation. The hemostasis valve of introducer set and the silica gel film at the hub is respectively used to stop blood stream and prevent seeping blood. It could avoid hemorrhage.
[Instruction for Use]
NOTE
Hold the guidewire anytime. According to the centimeter mark, adjust the insertion depth.
If widening percutaneous puncture is needed, do the operation with scalpel, its blade opposite to the guidewire.
Type I (Radial) | |||||||
Ref. NO. | Fr Size | Sheath | Dilator | Guide Wire | Needle (G/mm) |
||
I. D. (mm) |
L (mm) |
I. D. (mm) |
L (mm) |
O. D. (inch) |
|||
TR01 | 5F | 1.9 | 110 | 0.68 | 168 | 0.021 | 21/45 |
TR02 | 6F | 2.1 | 110 |
Characteristic:
Type I (Radial)
1.Longer taper ensure smooth transition between sheath and dilator
2.Excellent pushability, flexibility and radiopacity
3.0.021" stainless steel guide wire, hydrophilic guide wire available upon request
Type II (Radial)
1.Longer taper ensure smooth transition between sheath and dilator
2.Excellent pushability, flexibility and radiopacity
3.Hydrophilic guide wire with cannula needle is available
4.Hydrophilic coated sheath available
Type III (Femoral)
1.Excellent pushability, flexibility and radiopacity
2.Smooth transition between sheath and dilator
3.No leakage under 80kpa
4.Hydrophilic coated sheath available